KROHNE IFC 110 Converter User Manual

Page 17

Advertising
background image

Part A System installation and start-up Sect. 2.2

05/2003

IFC 110 F

17

2.2

Current output I


• The current output is electrically isolated from all other circuits.


• Setting data and functions can note down in the Table in Sect. 3.

Please also refer to Section 2.7 ”Standard factory settings”.


• All operating data and functions are adjustable (see Sections 4.4 and 5.6, Fct. 1.05).


• Max. load :

active operation 15-500

Ω

passive operation

≤ 800 Ω


• Selfcheck : - interrupting the mA loop, and

- short-circuit of mA loop via test function, see Fct. 2.03

or when power supply is switched on in Fct. 3.07

Error message on display (see Fct. 1.04, Section 5.4) and/or

status output (see Fct. 1.07-1.10, Section 5.8).


• Current value for error identification is adjustable, see Fct. 1.05 and Section 5.6.


• Range change-over, automatically or externally by control input,

see Sections 4.4 and 5.19, Fct. 1.07-1.10 and 1.11-12.

Setting range from 5-80% of Q

100%

(corresponding low to high range ratio from 1:20 to 1:1.25).

Change-over from high to low range at approx. 85% of low range and vice versa at approx.

98% of low range.

The active range is signalled via one of the four status outputs.

• Forward/reverse flow measurement (F/R mode) is possible (see Section 5.15).


• Connection diagrams see Section 2.6.