KROHNE IFC 110 Converter User Manual

Page 18

Advertising
background image

Sect. 2.3.1

Part A System installation and start-up

18

IFC 110 F

05/2003

2.3

Pulse outputs P and A1


2.3.1 Pulse output P for electronic totalizers (EC)

• Pulse output P is electrically isolated from all other circuits.


• Setting data and functions can note down in the Table in Sect. 3.

Please also refer to Section 2.7 ”Standard factory settings”.


• All operating data and functions are adjustable, see Sections 4.4 and 5.7, Fct. 1.05.


• Active mode:

uses the internal power supply, terminals E+/E-

Passive mode:

requires external power supply, U

ext

≤ 32V DC/24V AC, I ≤ 30mA


• Max. adjustable frequency 10 kHz


• Scaling

in pulses per unit time (e.g. 1000 pulses/s at Q

100%

flow) or

in pulses per unit volume (e.g. 100 pulses/m³ or US Gal).


• Pulse width

symmetric, pulse duty factor 1:1, independent of output frequency,

automatic, with optimum pulse width,

pulse duty factor approx. 1:1 at Q

100%

, or

pulse width range from 0.01 to 1 s adjustable as required for

correspondingly lower output frequency.


• Forward/reverse flow measurement (F/R mode) is possible, see Section 5.15.


• Connection diagrams see Section 2.6


• Schematic wiring diagram for pulse output P for electronic totalizers EC

Similar to a relay contact, this pulse output switches direct and alternating voltages.