KROHNE IFC 110 Converter User Manual

Page 35

Advertising
background image

Part B IFC 110 F Signal converter Sect. 4.4

05/2003

IFC 110 F

35

Fct. Text

Description

and

setting

1.07 STATUS

A1

Status

output

A1

(terminal A1 connected as status output A1 or

as a 2nd pulse output A1, see Fct. 3.07 HARDWARE, ”terminal A1”)

1.08 STATUS

A2

Status

output

A2

1.09

STATUS D1

Status output D1

1.10

STATUS D2

Status output D2

→

● OFF

● ON

● ALL ERROR

● FATAL.ERROR

● INVERS D1 (inverse mode of D1 and D2)

● INVERS A1 (inverse mode of A1 and A2 possible only if A1

is operated as status output, see Fct. 3.07 HARDWARE, ”terminal A1”)

● SIGN I, P or P2

(F/R

mode)

● OVERFL. I, P or P2

(overloading

the

outputs)

}

dynamic behaviour
of outputs see Fct. 1.02
TIMECONST.:
I = ONLY I
P or P2 = ALL

● EMPTY PIPE (”tube empty” signal only with built-in option)

● TRIP. POINT

Press → key to change to character.

Selection:

● + DIR.

● - DIR.

● 2 DIR.

Press ↵ key to change to numerical setting.

Setting

range: 000 - 150 PERCENT

● AUTO. RNG. Setting range: 05-80 PERCENT (= lower to upper range ratio

1:20 to 1:1.25, value must be higher than that of Fct. 1.03 L.F. CUTOFF)

Fct. 1.07 to 1.1

0

are configured in

accordance

w

ith the same setting

mode. Functions

set for one of th

e

status outputs ar

e no longer available

for the oth

er stat

us outputs.

Press ↵ key to change to numerical setting.

Press ↵ key to return to Fct. 1.06, 1.07, 1.08 or 1.09.

Fct.

Text

Description and setting

1.11

CONTROL C1

Control input C1 and C2

1.12 CONTROL

C2 ● OFF

● EXT. RNG. (external range change)

→

Setting

range: 05-80 PERCENT (= lower to upper range ratio 1:20 to 1:1.25,

value must be higher than that of Fct. 1.03 L.F. CUTOFF)

Press ↵ key to change to numerical setting.

● OUTP. HOLD (hold output values)

● OUTP. ZERO (set outputs to ”min. values”)

● TOTAL.RESET (reset the totalizer)

● ERROR.RESET (delete error messages)

Press ↵ key to return to Fct. 1.11 or 1.12 CONTROL C1 or C2

Fct.

Text

Description and setting

2.0

TEST

Test menu

TEST Q

Test measuring range Q

Precautionary

query

● SURE NO

Press

↵ key to return to Fct. 2.01 “TEST Q”.

● SURE YES

Press

↵ key, then use ↑ key to

select value: -110 / -100 / -50 / -10 / 0 / +10 / +50 / +100 / +110 PCT.

of set full-scale range Q

100%

.

Displayed value is available at outputs I and P.

Press

↵ key to return to Fct. 2.01 “TEST Q”.

2.02

HARDW. INFO

Hardware information and error status

Before consulting factory, please note down all 6 codes.

→ MODUL ADC

X . X X X X X . X X

Y Y Y Y Y Y Y Y Y Y

Press

↵ key to transfer to “MODUL IO“.

→ MODUL IO

X . X X X X X . X X

Y Y Y Y Y Y Y Y Y Y

Press

↵ key to transfer to “MODUL DISP.“.

→ MODUL DISP.

X . X X X X X . X X

Press

↵ key to return to

Y Y Y Y Y Y Y Y Y Y

Fct. 2.02 “HARDW. INFO“.

2.03

HARDW. TEST

Hardware test (Precautionary query)

- SURE NO

Press

↵ key to return to Fct. 2.03 ”HARDW. TEST”.

-

SURE

YES Press ↵ key to start test, duration approx. 60 s

If errors are found, the first one is displayed. Press

↓ key to

display next error. List of errors see Section 4.5.

Press

↵ key to return to Fct. 2.03 ”HARDW. TEST”.