1.3.2 Stripping (preparation) of signal cables

Please note: The numbers in the drawings designate the stranded drain wires of signalling cables
A and B, see sectional drawings in Sect. 1.3.1.

Flow sensor

Converter

Length flow

sensor

Length Converter

mm

inch

mm

inch

a 90

3.60

a 50

2.00

b

8

0.30

b

8

0.40

c

25

1.00

d

8

0.40

d

8

0.30

e

20

0.80

e

70

2.80


Signal cable A (type DS),

double shielding

for flow sensor

Signal cable A (type DS),

double shielding

for IFC 110 F Converter


Signal cable B (type BTS),

with triple shielding (bootstrap)

for flow sensor

Signal cable B (type BTS),

with triple shielding (bootstrap)

for IFC 110 F Converter




Customer-supplied materials

W

Insulation tubing (PVC), Ø 2.0-2.5 mm (Ø 1”)

X

Heat-shrinkable tubing or cable sleeve

Y

Wire end sleeve to DIN 41 228: E 1.5-8

Z

Wire end sleeve to DIN 41 228: E 0.5-8



Signal cable A

bending radius
≥ 50 mm (≥ 2”)

Signal cable A

bending radius
≥ 50 mm (≥ 2”)

Signal cable B

bending radius
≥ 50 mm (≥ 2”)

Signal cable B

bending radius
≥ 50 mm (≥ 2”)