13 flow sensor – KROHNE IFC 110 Converter User Manual

Page 49

Advertising
background image

Part B IFC 110 F Signal converter Sect. 5.13

05/2003

IFC 110 F

49

5.13 Flow

sensor

Fct. 3.02 FLOW METER
Press

→ key

→ DIAMETER = set meter size (see instrument nameplate), press → key

Select size from table of meter sizes:
• DN 2.5 – 1200 equivalent to

1

/

10

- 48 inch

• DN 1300 – 3000 equivalent to 52 - 120 inch, see Section 8.6.

Select with

↑ and ↓ keys.

Press

↵ key to change to subfunction ”FULL SCALE”.

→ FULL SCALE = set full-scale range, press → key

Set as described in Section 5.1.

Press

↵ key to change to subfunction ”GK VALUE”.

Please note that if ”VALUE P” or ”VALUE P2” is displayed after pressing

↵ key:

PULSE/VOL. is set in Fct. 1.06 PULS P and/or in Fct. 1.07 PULS 2 A1, subfunction ”SELECT P”
and/or ”SELECT P2”. Due to the changed full-scale range Q

100%

, the output frequency (F) of the

pulse outputs is either exceeded or not reached:

P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

100%

Change pulse value accordingly, see Section 5.08 Pulse output P, Fct. 1.06
and/or 2nd pulse output A1, Fct. 1.07.

→ GK VALUE = set flow sensor constant GK, press → key

• 1.0000 - 9.9999 (note information on instrument nameplate, do not change setting)

Change flashing number (cursor) with keys

↑ and ↓ .

Use

→ and ← keys to shift cursor 1 place to right or left.

Flashing numbers (cursor) can also be directly set with the 10-key keyboard.
Press

↵ key to change to subfunction ”FIELD FREQ.”

→ FIELD FREQ. = set magnetic field frequency, press → key

• 1/2

• 1/6

(1/2, 1/6, 1/18 or 1/36 of power frequency, see instrument nameplate,

• 1/18

• 1/36 do not change setting, exceptions see Sections 6.4-6.10)

Select with

↑ and ↓ keys.

Press

↵ key to change to subfunction ”FLOW DIR.”

(for DC instruments change to subfunction ”LINE FREQ.”).

→ LINE FREQ. = set power frequency customary in country

where instrument is used, press

→ key

(Please note: only applies to instruments with DC power unit)
• 50 Hz

Select with

↑ and ↓ keys.

• 60 Hz

Press

↵ key to change to subfunction ”FLOW DIR.”.

→ FLOW DIR. = set flow direction, press → key

• + DIR.

(for identification of flow direction see ”+” arrow on flow sensor;

• - DIR.

F/R mode: identification of ”positive” flow direction)

Select with

↑ and ↓ keys.

Press

↵ key to return to Fct. 3.02 FLOW METER.

Zero check, see Fct. 3.03 and Section 7.1.

Please refer to Section 2.7 ”Factory settings”.