7 standard factory settings – KROHNE IFC 110 Converter User Manual

Page 25

Advertising
background image

Part A System installation and start-up Sect. 2.7

05/2003

IFC 110 F

25

2.7

Standard factory settings


• All operating data are set at the factory in accordance with the specifications contained in the

order.

• If no specifications are made in the order, instruments will be delivered with the standard

parameters and functions indicated in the table below.

• To facilitate the start-up of the instrument, current and pulse outputs are set to handle

measurements in ”two flow directions” so that the current flow rates and volumes are displayed
and/or counted independent of the direction of flow. The figures displayed may have a
preceding sign.

• Such factory setting of current and pulse outputs may lead to measuring errors, particularly

when volumes are metered and totalized.

• If e.g. pumps are switched off and ”backflows” occur which are not within the low-flow cutoff

(SMU) range, or if separate displays and counts are required for both flow directions.

• To avoid faulty measurements, it may therefore be necessary to change the setting of the

following functions:

- SMU low-flow cutoff

Fct. 1.03, Section 5.3

- display

Fct. 1.04, Section 5.5

- current output I

Fct. 1.05, Section 5.6

- pulse output P

Fct. 1.06, Section 5.7

• For special applications, such as pulsating flows, see Sections 6.5 to 6.10


Standard factory settings

Fct. No. Function

Setting

Fct. No. Function

Setting

1.01

Full-scale range

See instr. nameplate 1.08

Status output A2 ON

of flow sensor

1.09

Status output D1 All error

1.02

Time constant

3 Sec. for display,

1.10

Status output D2 Indication F/R

pulse, current and

1.11

Control input C1 Totalizer reset

status ouputs

1.12

Control input C2 OFF

1.03

Low-flow cutoff

ON: 1%

3.01

Language

German

OFF: 2%

3.02

Flow sensor

1.04 Display

meter size

See instr. nameplate

flow rate

m³/h

direction of flow

+ direction,

totalizer m³

see

arrow

on

1.05 Current

output

I

active

flow

sensor

function

2 directions

3.04

Entry code

NO

range

4-20 mA

3.05

User unit

Liter/h

error detection

22 mA

3.06

Application

1.06

Pulse output P

flow

steady

function

2 directions

empty pipe

NO

pulse value

1000 pulses/Sec.

ADC gain

automatic

pulse width

symmetric

special filter

OFF

1.07

Pulse output 2, A1

3.07

Hardware

function 2

directions

terminal A1

pulse output A1

pulse value

1 pulse/s

selfcheck

NO

pulse width

50 ms