KROHNE IFC 110 Converter User Manual

Page 52

Advertising
background image

Sect. 5.17

Part B IFC 110 F Signal converter

52

IFC 110 F

05/2003

5.17 Applications


Fct. 3.06 APPLICAT.
Press

key twice


Set flow characteristics, select with

↑ or ↓ keys

STEADY

(steady flow)

PULSATING

(pulsating flow, e,g, caused by reciprocating pumps,
refer to Sections 6.5 to 6.10 ”Special applications”)


Press

↵ key to change to subfunction ”ADC GAIN”.


Set ADC GAIN, select with

orkeys

AUTO

(for homogeneous process liquids, low pulsation)

10

(for high solids contents or extremely pulsating flows)

30

(for solids contents or pulsating flows)

100

(high resolution even at low flows)


Press

key three times to return to Fct. APPLICAT.


Do not change the settings of subfunctions ”SPEC. FILT.”, ”LIMIT VAL.” and ”LIMIT CNT.” as
these functions are needed to obtain steady signals for display and outputs for special
applications, see Section 6.6.

5.18 Hardware

settings


Fct. 3.07 HARDWARE
Press

key


Define function of terminal A1, press

key

PULSOUTP.

(= pulse output)

Select with

orkeys,

STATUSOUTP.

(= status output)

}

press key to change to ”Selfcheck”


Carry out selfcheck during measurement? Press

key

NO

YES

Select with

orkeys, presskey to change to ”Field current”.


What is checked?

ADC gain and other parameters are continuously checked for their
permissible values and deviations.

Errors are only displayed when ”YES” is selected in Fct. 1.04 DISPLAY, subfunction ”DISP.
MSG.”. After acknowledging/deleting the errors in the ERROR/QUIT menu (see Section 4.6), the
tests described in a) and b) above are re-started. Test duration 4 to 20 minutes.

Select field current supply, press

key

INTERNAL

(DN 2.5–1600 / 1/10”–64”)

EXTERNAL

(see Section 8.6)

}

Select with ↑ or ↓ keys,


Press

key to return to Fct, 3.07 HARDWARE.