KROHNE IFC 110 Converter User Manual

Page 45

Advertising
background image

Part B IFC 110 F Signal converter Sect. 5.8

05/2003

IFC 110 F

45


PULSWIDTH = select pulse width, press key

AUTO

(automatic = 50 % of cycle duration of 100 % output frequency)

SYM.

(symmetric = pulse duty factor 1:1 across entire range)

SEC.

(variable) setting range 0.01 - 1.00 SEC

Select with

and keys.

Press

key to change to numerical setting.

Change flashing number (cursor) with keys

and ↓ . Useandkeys to shift cursor

1 place to right or left.
Flashing numbers (cursor) can also be directly set with the 10-key keyboard.

Press

key to change to subfunction ”VALUE P” and/or ”VALUE P2”.


VALUE P = set pulse value per unit volume

only appears when ”PULSE/VOL.” is selected in ”SELECT P”, press

key

XXXX PulS/m³

XXXX PulS/litre

XXXX PulS/US.Gal

XXXX PulS/user-defined unit, factory setting = ”Liter” or ”US MGal/day”, see Section 5.14

Select with

and keys.

Press

key to change to numerical setting, 1st number (cursor) flashes.

Set numerical value
XXXX

(setting range depends on pulse width and full-scale range:

P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

100%

)

Change flashing number (cursor) with keys

and ↓ . Useandkeys to shift cursor

1 place to right or left.
Flashing numbers (cursor) can also be directly set with the 10-key keyboard.
Press

key to return to Fct. 1.06 PULS P or to Fct. 1.07 PULS2 A1.

or

VALUE P = set pulse value per unit time

only appears when ”PULSE/TIME” is selected in ”SELECT P”, press

key

XXXX PulSe/Sec

XXXX PulSe/min

XXXX PulSe/h

XXXX PulSe/user-defined unit, factory setting = ”h” or ”day”, see Section 5.14

Select with

and keys.

Press

key to change to numerical setting, 1st number (cursor) flashes.

Set numerical value
XXXX

(setting range depends on pulse width)

Change flashing number (cursor) with keys

and ↓ . Useandkeys to shift cursor

1 place to right or left.
Flashing numbers (cursor) can also be directly set with the 10-key keyboard.
Press

key to return to Fct. 1.06 PULS P or to Fct. 1.07 PULS2 A1.


Please refer to Section 2.7 ”Factory settings”.

Refer to Section 2.6 for connection diagrams and to Section 5.16 for characteristics.

Advertising