KROHNE IFC 110 Converter User Manual

Page 70

Advertising
background image

Sect. 7.5

Part C Special applications, functional checks, service, and order numbers

70

IFC 110 F

05/2003

7.5

Faults and symptoms during start-up and flow measurement


• Most faults and symptoms occurring with the flowmeters can be eliminated by following the

instructions indicated in the following tables.


• For greater clarity, faults and symptoms in the tables are divided into different groups.


LED light-emitting diodes on the front panel (status messages)

D

display

I

current output I

P

pulse outputs P and A1

S

status outputs D1, D2, A1 and A2

C

control inputs C1 and C2


Before contacting the KROHNE Service Department,

please read the instructions in the table. THANK YOU.


Group LED

Display

Cause

Remedial action

LED 1

Both LEDs flash

A/D converter range

Reduce flow rate;

exceeded

if unsuccessful, test as

described

in

Section

7.6

Measuring tube drained,-

Fill measuring tube

A/D conv. range exceeded

LED 2

Red LED flashes

Fatal error, hardware and/or

Replace signal converter,

software fault

see Section 8.3

LED 3

Cyclic flashing of red

Hardware fault, watchdog

Replace signal converter,

LED, approx. 1 sec.

trips

see Section 8.3

LED 4

Red LED on

Hardware fault

Replace signal converter,

continuously

see

Section

8.3