Part B IFC 110 F Signal converter Sect. 5.3

05/2003

IFC 110 F

41

5.3 Low-flow

cutoff

SMU

Fct. 1.03 L.F. CUTOFF
Press

→ key


Select
• OFF

(fixed tripping points:

ON = 0.1 % / OFF = 0.2 %)

• PERCENT

(variable tripping points:

ON = 1 - 19 % / OFF = 2 - 20 %)

Select with keys

↑ and ↓ (only if PERCENT is selected).

1st number (cursor) flashes.

Setting the numerical value when ”PERCENT” is selected
• 01 to 19

(cutoff ”ON” value, left of hyphen)

• 02 to 20

(cutoff ”OFF” value, right of hyphen)

Change flashing number (cursor) with keys

↑ and ↓ .

Use

→ and ← keys to shift cursor 1 place to right or left.

Flashing numbers (cursor) can also be directly set with the 10-key keyboard.
Press

↵ key to return to Fct. 1.03 L.F. CUTOFF.

Note: the cutoff ”OFF” value must be greater than the cutoff ”ON” value.

5.4 Display

Fct. 1.04 DISPLAY
Press

→ key

→ DISP. FLOW = select unit for display of flow rate, press → key

• NO DISP

(no display)

• m³/h

(cubic metres per hour)

• Liter/Sec

(litres per second)

• US.Gal/min

(US gallons per minute)

• User-defined unit, factory setting = ”Liter/h” (litres per hour) or ”US MGal/day”,

see Section 5.14

• PERCENT

(percentage display)

• BARGRAPH

(numerical value and bar graph display in %)

Select with

↑ and ↓ keys.

Press

↵ key to change to subfunction ”DISP. TOTAL.”.

→ DISP. TOTAL. = select unit for totalizer display, press → key

• NO DISP.

(no display)

• OFF

(internal totalizer switched off)

• + TOTAL.

• – TOTAL.

• +/– TOTAL.

• SUM. (Σ)

• ALL (sequential)

Select with

↑ and ↓ keys.

Press

↵ key to change to display unit setting.


• m³

(cubic metres)

• Liter

(litres)

• US.Gal

(US gallons)

• User-defined unit, factory setting = ”Liter” (litres) or ”US MGal/day”, see Section 5.14

Select with

↑ and ↓ keys.

Use

→ key to change to totalizer format setting.