KROHNE IFC 110 Converter User Manual

Page 54

Advertising
background image

Sect. 6.4

Part B IFC 110 F Signal converter

54

IFC 110 F

05/2003

5.20 Range

change


Automatic range change by status output


Fct. 1.07 - 1.10 Status outputs A1, A2, D1 or D2
(Define operating mode of output terminal A1, see Section 5.18)

Press

→ key


Press

↑ key as often as required to set one of the status outputs to automatic range change

”AUTO. RNG.”

Press

↵ key to change to numerical setting, 1st number (cursor) flashes.

Change flashing number (cursor) with keys

↑ and ↓. Use → and ← keys to shift cursor

1 place to right or left.
Flashing numbers (cursor) can also be directly set with the 10-key keyboard.

Setting range: 5 – 80 PERCENT of Q

100%

(= low to high range ratio 1:20 to 1:1.25)


Press

↵ key to return to Fct. 1.07 - 1.10, status outputs A1, A2, D1 or D2.




External range change by control input


Fct. 1.11 or 1.12 Control inputs C1 or C2

Press

→ key


Press

↑ key as often as required to set one of the control inputs C1 or C2 to range change

”EXT. RNG.”

Press

↵ key to change to numerical setting, 1st number (cursor) flashes.

Change flashing number (cursor) with keys

↑ and ↓. Use → and ← keys to shift cursor

1 place to right or left.
Flashing numbers (cursor) can also be directly set with the 10-key keyboard.

Setting range: 5 – 80 PERCENT of Q

100%

(= low to high range ratio 1:20 to 1:1.25)


Press

↵ key to return to Fct. 1.11 or 1.12, control inputs C1 or C2.