Sect. 2.4

Part A System installation and start-up

20

IFC 110 F

05/2003

2.4

Status outputs A1 / A2 / D1 / D2

PLEASE NOTE:
The output terminal A1 can be used as status output A1 or as a 2nd pulse output A1 for
electromechanical totalizers.
Setting is as described in Fct. 3.07 HARDWARE, see Sections 4.4 and 5.18.

• Status outputs A1/A2 and D1/D2 with the common centre grounding contacts A⊥ and B⊥ are

electrically isolated from each other and from all other circuits.

• Setting data and functions can note down in the Table in Sect. 3.

Please also refer to Section 2.7 ”Standard factory settings”.

• All operating data and functions are adjustable, see Sections 4.4 and 5.8, Fct. 1.07-1.10.

• Active mode:

uses the internal power supply, terminals E+/E-

Passive mode:

requires external power supply, U

ext

≤ 32V DC/24V AC, I ≤ 100mA

(I

≤ 200mA for A1 in case of polarized DC operation, see Section 6.3)

• The following operating conditions can be signalled using the status outputs:

- flow direction (F/R mode)

- limits

- error messages

- active range in case of range change-over

- inverse operation of A1 and A2 or D1 and D2,

i.e. used as change-over switch with common centre grounding contact A

⊥ or D⊥.

• Connection diagrams see Section 2.6

• Schematic wiring diagram for status outputs A1/A2 and D1/D2.

This status outputs have MOSFET switches as outputs which switch direct and alternating
voltages similar to relay contacts.

2.5

Control inputs C1 and C2

• Control inputs C1 and C2 are electrically connected (common centre grounding

contact C

⊥) but electrically isolated from all other circuits.

• Setting data and functions can note down in the Table in Sect. 3.

Please also refer to Section 2.7 ”Standard factory settings”.

• All operating data and functions are adjustable, see Sections 4.4 and 5.9, Fct. 1.11-1.12.

• Active mode:

uses the internal power supply, terminals E+/E-

Passive

mode: requires external power supply, U

ext

≤ 32V DC/24V AC, I ≤ 10mA

• The following operating conditions can be initiated using the control inputs:

- external range change

- holding of output values

- zeroing the outputs

- resetting the internal totalizer

- resetting (deleting) the error messages

• Connection diagrams see Section 2.6