2 accuracy – KYORITSU 1062 User Manual

Page 12

Advertising
background image

70

3.2 Accuracy

Test conditions:
Temperature and humidity: 23±5ºC at 80%RH or less
Accuracy: ±(% of reading + digits)

Note: Each response time is a value to rated accuracy within selected range.

DC Voltage Measurement

V

Range

Resolution

Accuracy

Input Resistance

Maximum

Input Voltage

KEW1061, KEW1062

50mV

0.001mV

0.05+10

Approx. 100MΩ

1000V DC

500mV

0.01mV

0.02+2

2400mV

0.1mV

5V

0.0001V

0.025+5

10MΩ

1000V rms AC

50V

0.001V

0.03+2

500V

0.01V

1000V

0.1V

NMRR: 80dB or more 50/60Hz ±0.1%

(70dB or more 50/60Hz ±0.1% when 50mV Range)

CMRR: 100dB or more 50/60Hz (Rs=1kΩ)
Response time: 0.3 sec max.

AC Voltage Measurement [RMS]

V

KEW1061

AC Coupling, Rms-value detection, Crest factor*: <3

Range Resolution

Accuracy

Input

Impe-

dance

Maximum

Input

Voltage

10 to 20Hz 20Hz to

1kHz

1k to

10kHz

10k to

20kHz

500mV

0.01mV

1.5+30 *1

0.7+30 *1

2+50 *2

11MΩ

<50pF

1000V

rms AC

1000V DC

5V 0.0001V

50V

0.001V

10MΩ

<50pF

500V

0.01V

1000V *

0.1V

*2

*2

3+30 *2

-

*: Crest factor <1.5 at 1000V range
Accuracy *1: At 5 to 100% of range, *2: At 10 to 100% of range
CMRR: 80dB or more DC to 60Hz (Rs=1kΩ)
Response time: 1 sec max.

Advertising
This manual is related to the following products: