KYORITSU 1062 User Manual

Page 51

Advertising
background image

109

Table 1. Input Signal for Calibration

Range

Input 1

Input 2

Unit

DC 50mV

0.000

50.000

mV

DC 500mV

500.00

-

mV

DC 2400mV

2000.0

-

mV

DC 5V

5.0000

-

V

DC 50V

50.000

-

V

DC 500V

500.00

-

V

DC 1000V

1000.0

-

V

AC 50mV*1

5.000

50.000

mV

AC 500mV

50.00

500.00

mV

AC 5V *2

0.5000

5.0000

V

AC 50V

5.000

50.000

V

AC 500V

50.00

500.00

V

AC 1000V

100.0

1000.0

V

500Ω

0.00

500.00

5kΩ

0.0000

5.0000

kΩ

50kΩ

0.000

50.000

kΩ

500kΩ

0.00

500.00

kΩ

5MΩ

0.0000

5.0000

MΩ

50MΩ

0.000

30.000

MΩ

Continuity Check ( )

0.0

500.0

DC 500 µA

0.00

500.00

µA

DC 5000 µA

0.0

5000.0

µA

DC 50mA

0.000

50.000

mA

DC 500mA

0.00

400.00

mA

DC 5A

0.0000

5.0000

A

DC 10A

0.000

10.000

A

AC 500 µA

50.00

500.00

µA

AC 5000 µA

500.0

5000.0

µA

AC 50mA

5.000

50.000

mA

AC 500mA

50.00

400.00

mA

AC 5A

0.5000

5.0000

A

AC 10A

1.000

10.000

A

*1: For model KEW1062.

*2: For model KEW1062. Calibration points added in 5 V AC range (setting, input value).

[RMS], Filter OFF, Input1 (0.50000), Input2 (5.0000)

[MEAN], Filter OFF, Input1 (0.50000), Input2 (5.0000)

[RMS], Filter ON, Input1 (0.5000) , Input2 (5.0000)

50Hz/60Hz

50Hz/60Hz