KYORITSU 1062 User Manual

Page 45

Advertising
background image

103

<Default setting of dBm measurement reference resistance>

Set a default value of the reference resistance during dBm measurement mode.
1) Display “db-r” by using the LIGHT key.

The reference value appears on the display.

HOLD

2) Select the reference resistance by using the ▲ (RANGE) key or ▼ (REL) key.

3) Press the HOLD key to save the setting.

“SEt” appears and then “db-r.”

Setting values of reference resistance

4, 8, 16, 32, 50, 75, 93, 110, 125, 135, 150, 200, 250, 300, 500, 600, 800, 900, 1000,

1200Ω(default value is 600Ω)

<Default setting of detection method during AC measurement> (KEW1062 only)

Set a default setting of detection methods during AC measurement.
RMS or MEAN: The default setting is RMS.
1) Display “Ac” by using the LIGHT key.

HOLD

: RMS

: MEAN

2) Select the detection method by using the ▲ (RANGE) key or ▼ (REL) key.

3) Press the HOLD key to save the setting.

4) “SEt” appears and then “Ac.”

<Default value of LOGGING interval>

Set a default value of the saving interval during LOGGING mode.

1) Display “LG. int” by using the LIGHT key.

The default setting is 1 sec.

HOLD

2) Select the saving interval by using the ▲ (RANGE) key or ▼ (REL) key.

3) Press the HOLD key to save the setting. “SEt” appears and then “LG. int.”

Settings of saving interval

1, 2, 5, 10, 30, 60, 600, 1800 sec