KYORITSU 1062 User Manual

Page 52

Advertising
background image

110

<Table 2>

After completing the calibration for ranges in Table 1, carry out calibration for “ Frequency
Characteristic”.
The calibration for is frequency characteristic required for AC voltage and AC current ranges
(marked with ).
Calibration is carried out at the designated frequency in table 2.

1) Set the calibrator to Input value as an input to the instrument.

2) Press the MEMORY key.

3) After 20 seconds, buzzer sounds and the instrument confirm calibration.

(Do not next key operation till buzzer sounds.)

Table 2. Input Signal for Calibration

Range

Input

Unit

AC 50mV*1

50.000

mV

AC 500mV

500.00

mV

AC 5V

5.0000

V

AC 50V

50.000

V

AC 500V

500.00

V

AC 1000V 600Hz

1000.0

V

AC 500μA

500.00

μA

AC 5000μA

5000.0

μA

5kHz

1kHz

*1: For model KEW1062.