KYORITSU 1062 User Manual

Page 18

Advertising
background image

76

Continuity Check

Maximum Reading 5000

Range

Resolution Range of Operation Measuring

Current

Open-loop

Voltage

Input Protective

Voltage

500Ω

0.1Ω

The buzzer turns on

for resistances lower

than 100±50Ω.

Approx.

0.5mA

<5V

1000V rms

Diode Test

Range

Resolution

Accuracy Measuring Current

(Vf=0.6V)

Open-loop

Voltage

Input Protective

Voltage

2.4V

0.0001V

1+2

Approx. 0.5mA

<5V

1000V rms

Temperature Measurement TEMP

Range

Resolution

Accuracy

Input Protective

Voltage

-200 to 1372°C

0.1°C

1%+1.5°C

1000V rms

-328 to 2501.6°F

0.1°F

1%+2.7°F

Use optional Temperature Probe: Thermocouple Type K

Capacitor Measurement

Maximum Reading 5000

Range

Resolution

Accuracy

Input Protective Voltage

5nF

0.001nF

1+5

*1

1000V rms

50nF

0.01nF

500nF

0.1nF

5μF

0.001μF

50μF

0.01μF

500μF

0.1μF

2+5

5mF

0.001mF

3+5

50mF

0.01mF

*1: Accuracy is specified after zero adjustment (capacitor).

Advertising