3. Specifications

3.1 General Specifications

Measurement function:

DC Voltage, AC Voltage, DC Current, AC Current, Resistance,
Frequency, Temperature, Capacitor, Duty cycle ratio,
Decibel (dBV, dBm), Continuity Check, Diode Test, LowPower-Ω*

Other functions:

Data Hold (D•H), Auto Hold (A•H), Peak Hold* (P•H),
Range Hold (R•H), Maximum value (MAX), Minimum value (MIN),
Average value (AVG), Zero Adjustment (Capacitor, Resistance),
Relative values, Save to Memory, LCD backlight.

LowPower-Ω: Measures resistance under lowmeasurement current.

*: For model KEW1062 only

Measuring method:

Σ modulation

Display:

5-digit (LCD)/7-segment

Maximum Reading: 50000

Polarity Indicator: “–” Appears automatically when the polarity is
negative

Overrange Indicator: “ OL ”

Low-battery Indicator: “

” Appears when the batteries become

low

Measurement cycle: 6 times per second

(except frequency measurement: one times per second, Resistance
measurement : four times per second, capacitor measurement (50mF):
max. 0.03 time per second)

Bar graph display 15 times per second

Operating temperature and humidity ranges:

-20 to 55ºC, 80%RH or less (no condensation)

70%RH or less at 40 to 55ºC.

Storage temperature and humidity ranges:

-40 to 70ºC, 70%RH or less (no condensation)

Temperature coefficient:

(Accuracy at 23±5ºC) × 0.05/ºC or less
At -20 to 18ºC and 28 to 55ºC
When continuously DCV and DCA measuring, add 1digit / ºC
(except for 50mV, 5A, 10A range, add 3digits / ºC).

Power supply:

AA-size (R6) 1.5V batteries: 4