5. User Calibration Function

It is recommended that the instrument be calibrated periodically.
The instrument can be calibrated.

CAUTION

To avoid electrical shock

• Only authorized engineers are allowed to calibrate the instrument

using dedicated facilities.

• Connect the calibrator to the instrument with the calibrator’s test leads.

• Before carrying out calibration, read the instruction manual of the calibrator.

• Temporarily remove the test leads from the instrument before switching measurement mode

(function).

<Conditions of calibration>

Calibrator: With accuracy higher than of this instrument

Ambient Environment:

Temperature: 23±3ºC

Humidity: 55%RH or less

Leave the instrument for 30 minutes under above conditions

before carrying out calibration.

After reference valve of Calibrator stabilizes, Press the key to confirm for calibration valve.