4.6 Set-up Function

The following settings can be made using the Set-up function:
• time setting

• default setting of dBm measurement reference resistance

• default setting of detection method during AC measurement

• default setting of LOGGING interval

• default setting of number display/time display during LOGGING mode

• sound on/off setting (beep of buzzer)

• reset to factory preset mode

1) Press the SHIFT key shows “Shift” on the display.

2) Press the LIGHT key while holding down the SHIFT key changes the mode to Set-up mode

(from Set-up to time display).

3) Press the LIGHT key changes the setting items accordingly.

4) Change values by using the ▲ (RANGE) key or ▼ (REL) key.

5) Press the HOLD key to save/finish each setting.

“SEt” appears and the display returns to the setting items.

6) Hold down the LIGHT key for more than one second to return from Set-up mode to

measurement mode.

Note

To cancel any setting, hold down the LIGHT key for more than one second, or turn off by
using the function key.

<Time setting>

Set the time to be displayed during LOGGING mode. Be sure to set the time after changing the
batteries.
1) Display “CLOCK” by using the LIGHT key.

The first two digits of the time blink on the sub-display.

HOLD

2) Set the present hour by using the ▲ (RANGE) key or ▼ (REL) key.

3) Press the LIGHT key to make the last two digits blink.

4) Set the present minute by using the ▲ (RANGE) key or ▼ (REL) key.

5) Press the HOLD key to save the setting.

“SEt” appears and then “CLOCK.”