6.2 Fuse Replacement

If a current greater than the rated value flows when the instrument is in the current-
measurement range, a protection fuse may blow.
If this happens, replace that fuse. The instrument contains the following types of fuses.

W ARNIN G

• Turn the function switch to OFF(turn off the power).

• Be sure to disconnect the instrument from the circuit under test and test leads before

replacing the fuses.

• Do not operate the instrument with the casing left open.

• In order to avoid damage to the instrument or any possible accident, use fuses of the

specified rating.

   Fuse rating: F1

M-8926 (440mA/1000V, SIBA GmbH & Co. KG, 50 210 06.0.44)

F2

M-8927 (10A/1000V, SIBA GmbH & Co. KG, 50 199 06.10)

High breaking capacity type

To replace the fuse:

1) Remove the screw on the back of the casing.

2) Remove the back cover.

3) Remove the blown fuse from the fuse holder.

4) Install a new fuse in the holder.

(Make sure the fuse rating.)

5) Close the casing and fasten it with the screw.

F1:

(440mA/1000V)

F2:

(10A/1000V)
Screw hole

<Rear View>

Back cover removed