KYORITSU 1062 User Manual

Page 16

Advertising
background image

74

AC Current Measurement [MEAN]

A

KEW1062

MEAN value detection, RMS value calibration (sine wave)

Range

Resolution

Accuracy

Voltage

Drop

Maximum Input

Current

10 to 20Hz 20 to 500Hz 500Hz to

1kHz

500μA

0.01μA

2+20

1.5 +20

2+30

<0.11

mV/μA 440mA Protected

by a 440mA/

1000V fuse.

5000μA

0.1μA

50mA

0.001mA

<4 mV/mA

500mA*3

0.01mA

5A

0.0001A

3+20

2+20

4+30

<0.1 V/A

10A Protected

by a 10A/

1000V fuse.

10A

0.001A

Accuracy At 5 to 100% of range, At 10 to 100% of range for 10A Range
*3: Maximum measurement current : 440mA at 500mA range
Response time: 1 sec max.

DCA+ACA

+

KEW1061

Maximum Reading 50000, Crest factor: <3

Range

Resolution

Accuracy

Voltage Drop

Maximum Input

Current

DC, 10 to

20Hz

DC, 20Hz to

1kHz

500μA

0.01μA

2+10

1.5+10

<0.11 mV/μA

440mA Protected by a

440mA/1000V fuse.

5000μA

0.1μA

50mA

0.001mA

<4mV/mA

500mA*3

0.01mA

5A

0.0001A

<0.1V/A

10A Protected by a

10A/1000V fuse.

10A

0.001A

Accuracy At 5 to 100% of range, At 10 to 100% of range for 10A Range
*3: Maximum measurement current : 440mA at 500mA range
Response time: 2 sec max.