Caution – KYORITSU 1062 User Manual

Page 33

Advertising
background image

91

4.3.12 DC, AC Current Dual Display ( •

)

1) Turn the function switch to the “µA,” “mA” or “A” position.

(If the magnitude of the current being measured is not known,

select the “A” position. Make sure the current being measured

is no more than 440 mA before the “µA” or “mA” position is

selected.)

2) Press the SELECT key three times to select DC/AC dual

display.

DC current measurement appears on the main display and AC

current on the sub-display.

3) Plug the black test lead into the “COM” input terminal and the

red test lead into the “A” input terminal.

If the current is in the order of mA or less, plug the red test lead

into the “μA/mA” input terminal.

4) Connect the test leads to the circuit under test and then read the

value when it stabilizes.

Test leads

Red

Black

Test leads

Red

Black

4.3.13 Capacitor Measurement (

)

CAUTION

To avoid damage to instrument

• Turn off the power to the circuit under test before starting measurement in order to prevent

any excessive voltage from being applied to the instrument.

• Before starting measurement, be sure to discharge the capacitor under check.

1) Turn the function switch to the “

”position.

2) Plug the test leads into the input terminals.

3) Open the test lead and press the REL key in 5nF range to adjust the

capacitance to zero. (The display shows “ 0.000 ”.)

4) Connect the test leads to the circuit under test and then read the value

when it stabilizes.

Test leads

Red

Black

Note

The value (zero adjustment) remains displayed until power-off.