75

DCA+ACA

+

KEW1062

Maximum Reading 50000, Crest factor: <3

Range Resolution

Accuracy

Voltage Drop Maximum Input

Current

DC, 10 to

20Hz

DC, 20Hz

to 1kHz

DC, 1k to

5kHz

500μA

0.01μA

1.5+10

1+10

1.5+10

<0.11 mV/μA 440mA Protected

by a 440mA/

1000V fuse.

5000μA

0.1μA

50mA

0.001mA

<4 mV/mA

500mA*3

0.01mA

5A

0.0001A

2+10

1.5+10

3+10

<0.1 V/A

10A Protected by a

10A/1000V fuse.

10A

0.001A

Accuracy At 5 to 100% of range, At 10 to 100% of range for 10A Range
*3: Maximum measurement current : 440mA at 500mA range
Response time: Approx. 2 sec

Resistance Measurement Ω

Range

Resolution

Accuracy

Maximum

Measuring

Current

Open-loop

Voltage

Input Protective

Voltage

KEW1061 KEW1062

500Ω

0.01Ω

0.1+2

*1

0.05+2

*1

<1mA

<2.5V

1000V rms

5kΩ

0.0001kΩ

<0.25mA

50kΩ

0.001kΩ

<25μA

500kΩ

0.01kΩ

<2.5μA

5MΩ 0.0001MΩ

0.5+2

<1.5μA

50MΩ

0.001MΩ

1+2

<0.13μA

*1: Accuracy is specified after zero adjustment (resistance).
Response time: 1 sec max. at 500Ω to 500kΩ

5 sec max. at 5MΩ to 50MΩ

LowPower-Ω LP-Ω

Maximum Reading 5000

Range

Resolution

Accuracy

Maximum

Measuring

Current

Open-loop

Voltage

Input Protective

Voltage

KEW1062 only

5kΩ

0.001kΩ

0.2+3

<10μA

<0.7V

1000V rms

50kΩ

0.01kΩ

<1.0μA

500kΩ

0.1kΩ

<0.6μA

5MΩ

0.001MΩ

1+3

<0.05μA

LowPower-Ω: Measures resistance under lowmeasurement current.