KYORITSU 1062 User Manual

Page 53

Advertising
background image

111

l

Calibration of Capacitor Range

Before start calibration of the Capacitor range,
turn the function switch back to the OFF position.
1) Turn the function switch from the OFF position to the

(Capacitor) position while

pressing the SELECT and RANGE keys at the same time.

The display shows the “ CAL ” symbol then the “ PASS ” symbol.

2) Press the SELECT key. (The display shows the “ - ” symbol.)

3) Press the HOLD key twice. (The display shows the “ - - - ” symbol.)

4) Press the RANGE key. (The display shows the “ nF ” symbol.)

5) Connect the instrument to the calibrator with the test leads.

6) Set the calibrator to Input 1 value as an input to the instrument.

7) Press the HOLD key to confirm.

8) Set the calibrator to Input 2 value as an input to the instrument.

9) Press the HOLD key to confirm.

10) Carry out calibration of other ranges by repeating steps 6) to 9).

11) To quit calibration, turn the function switch back to the OFF position.

Range

Input 1

Input 2

Unit

5nF

0.500

5.000

nF

50nF

5.00

50.00

nF

500nF

50.0

500.0

nF

5μF

0.500

5.000

μF

50μF

5.00

50.00

μF

500μF

50.0

500.0

μF

5mF

0.500

5.000

mF

50mF

5.00

40.00

mF