Specifications, Table 9: general specifications, Table 10: dc voltage measurement/source – Martel Electronics DMC1410 User Manual

Page 100: Table 11: dc ma measurement/source

Advertising
background image

10. Specifications

One year at 23°C ± 5°C, unless specified otherwise. Outside of this range
the stability of the measurements is ± 0.003% of FS/°C.

Table 9: General Specifications

Operating Temperature

-10°C to 50°C

Storage Temperature

-20°C to 60°C

Power

4 X AA batteries; Alkaline or optional rechargeable

Battery Life

Approximately 20 hours

Serial Communications ASCII; RS232 or USB depending on the communications

cable used

CE - EMC

EN61326: 2006, ANNEX B Criterion A

Safety

CSA C22.2 No. 1010.1: 1992

Size (with holster)

8.5” x 4” x 2” (21.6cm x 10.2cm x 5.1cm)

Weight

Approximately 28oz. (0.8kg)

Table 10: DC Voltage Measurement/Source

Range

Accuracy

(% of reading ± floor)

Read: Isolated(Upper Display)

0.000V - 30.000V

0.01% ± 2mV

Read: non-Isolated(Lower Display)

0.000V - 20.000V

0.01% ± 2mV

Source

0.000V - 20.000V

0.01% ± 2mV

Maximum current output in voltage ranges is 3mA with an output impedance of <= 1Ω.

Table 11: DC mA Measurement/Source

Range

Accuracy

(% of reading ± floor)

Read: Isolated(Upper Display)

0.000mA - 24.000mA

0.01% ± 2µA

Read: non-Isolated(Lower Display)

0.000mA - 24.000mA

0.01% ± 2µA

Source

0.000mA - 24.000mA

0.01% ± 2µA


Maximum load on mA source is 1000Ω. Voltage input range on simulate mode 5V - 30V.

98