Martel Electronics DMC1410 User Manual

Page 85

Advertising
background image

The letter U indicates an uncalibrated downloaded tag, the
remainder have been calibrated. Tag numbers are displayed on
multiple pages, six tags to a page. Press the UP ARROW and
DOWN ARROW keys to scroll the tag selector highlight one tag at a
time, scrolling from page to page, cycling from top to bottom of the
list and vice versa. Use the LEFT ARROW and RIGHT ARROW keys
to scroll the tag selector highlight one page at a time, cycling from
top to bottom of the list and vice versa.
Press the EXIT function key to return to the review selection menu
without selecting a tag.
When the desired tag number is highlighted, press the SELECT
function key to display the first page of data showing the tag number,
calibration date and time, and the first few identification fields.

Press the UP ARROW and DOWN ARROW keys to scroll through the
remaining pages of data, cycling from the last page to the first page
and vice versa. The second page contains the remaining
identification fields and the validation error tolerance. The third page
contains the instrument input setup. The fourth page contains the
instrument output setup. The fifth and subsequent pages display the
test results. On these pages only, use the LEFT ARROW and RIGHT
ARROW keys to switch between the As Found results and the As Left
results.
Note that the input/output orientation of the data on these pages is
with respect to the instrument under test. This is in contrast to the

83