Martel Electronics DMC1410 User Manual

Page 73

Advertising
background image

If the data displayed is correct, press the SAVE function key to save
the data and step to the next point. The point number in the upper
right increases to show that the test data is saved. If the pulse train
or manual data must be corrected before saving it, press the LEFT
ARROW key to change back to the TRIG/READY prompts and repeat
the test.
Once the desired number of points have been tested, press the
DONE function key to end the data collection stage and proceed to
the entry of test identification data.
If for any reason the data collection is to be terminated without
saving any data, press either the HOME key or the ABORT function
key. A prompt is displayed requesting confirmation that the test data
accumulated so far is to be abandoned. Press the NO function key
to return to the data collection process at the point where it was
interrupted. Press the YES function key to abandon the data
collected.

Manual Input/Output Type Prompt
When the input/output combination includes a manual input, manual
output, or both, a prompt similar to the following is displayed for
each test point.

The current test type, AF for As Found, and point number are shown
in the upper right corner of the display. Note that the point number
is the next point to be tested and data has not been saved for it yet.
If the calibrator output is pressure, adjust the external pressure
source until the displayed output value matches the desired test
value. If the calibrator output is manual, adjust the external source to
the desired test value. For all other output types, enter the output
value to be generated using the numeric keys, followed by ENTER.

71

Advertising