Martel Electronics DMC1410 User Manual

Page 51

Advertising
background image

two digit hour in 24 hour format (0 to 23), nn is the two digit minute
(0 to 59), and ss is the two digit second (0 to 59). Note that all two
digit values must have a leading zero if they have a value less than
10. Returns <Complete> if successful, otherwise an error message
enclosed in angle brackets. For example:

SET_CLOCK 2006 03 20 09 16 33

SIM
Sets the output for current simulation. This command also switches
the calibrator into mA simulation mode. A number and a unit must
be entered after the command. For example:

SIM 5 MA sets the current simulation at 5 mA

SIM?
Returns the output of the current simulation. With the example
above, the output would be:

5.000000E-03, A

TAG_CLEAR tag
Clears test data for the specified tag from the calibrator. Once the
test data is cleared, it can not be recovered. Returns <Complete> if
successful, otherwise an error message enclosed in angle brackets.
For example:

TAG_CLEAR FT-8567

TAG_CLEAR_ALL
Clears test data for all tags from the calibrator. Once the test data is
cleared, it can not be recovered. Returns <Complete> if successful,
otherwise an error message enclosed in angle brackets.

TAG_DNLD
See the separate Tag Download and Upload Procedures manual for
detailed instructions on using this command to download tag
calibration test configuration data from a PC to the calibrator.

TAG_UPLD
See the separate Tag Download and Upload Procedures manual for
detailed instructions on using this command to upload tag
calibration test results to a PC from the calibrator.