46

FREQ_UNIT
Sets the unit for frequency. There are three ranges of frequencies for
frequency and pulse modes, CPM (cycles per minute), Hz, and kHz.
Use this command to select the right range. For example:

FREQ_UNIT HZ sets the frequency to Hz range


FREQ_UNIT?
Returns the frequency unit currently being used by the frequency
and pulse modes.

FUNC?
Returns the current mode of the upper and lower displays. For
example if the calibrator is set to volts on the upper display, and
pressure on the lower display, FUNC? Would return:

DCV, PRESSURE

GET_CLOCK
Returns the current calibrator time and date as "yyyy/mm/dd
hh:mm:ss", for example:

2006/03/25 19:02:56

GET_SN
Returns the calibrator serial number of up to 10 digits, for example:

12345678

LOCAL
Restores the calibrator to local operation if it was in remote mode.
Also clears LOCKOUT if the unit was in lockout mode.

LOWER_MEAS
Sets the lower display to measure mode. The command is followed
by any of the parameters except for pulse and mA sim, which are
source only. Enter DCI for mA, DCV for volts, TC for thermocouple,
RTD for RTD, FREQUENCY for frequency, and PRESSURE for
pressure. For example:

LOWER_MEAS DCV sets the lower display mode to VOLTS IN