24

2. Enter the RTD custom setup main menu, and select [SET

CUSTOM].

3. Using the keypad, enter the values that the calibrator prompts

for: minimum temperature, maximum temperature, R0, and
the values for each of the temperature coefficients.

The custom function uses the Calendar-Van Dusen equation for
outputting and measuring custom RTDs. The coefficient C is
only used for temperatures below 0°C. Only A and B coefficients
are needed for the range above 0°C, so coefficient C should be
set to 0. The R0 is the resistance of the probe at 0°C. The
coefficients for PT385, PT3926, and PT3616 are shown in Table 4
below.

Table 4. RTD Coefficients

RTD

Range(°C) R0

Coefficient A Coefficient B Coefficient C

PT385

-260 - 0

100

3.9083x10-3

-5.775x10-7

-4.183x10-12

PT385

0 - 630

100

3.9083x10-3

-5.775x10-7

---

PT3926

Below 0

100

3.9848x10-3

-5.87x10-7

-4x10-12

PT3926

Above 0

100

3.9848x10-3

-5.87x10-7

---

PT3916

Below 0

100

3.9692x10-3

-5.8495x10-7

-4.2325x10-12

PT3916

Above 0

100

3.9692x10-3

-5.8495x10-7

---

5. Using Isolated Measure Modes (Upper

Display)

5.1 Measuring volts and mA

Use the following steps to measure the voltage or current output of a
transmitter.
1. Select the upper display from the Main Menu.
2. Select the desired primary parameter to be measured. Connect

the leads to the isolated inputs of the calibrator, as in Figure 16.