L_PRES_UNIT
Sets the unit for measuring pressure on the lower display. Add the
unit after the command. The available pressure units and their syntax
are shown in the Table 7. (Parameter Units).
For example:

L_PRES_UNIT KPAL sets the pressure unit to kiloPascals

OUT
Sets the output of the calibrator. This command may be used to
output mA, volts, frequency, temperature, and ohms. Frequency
output, which is set by the command FREQ_TYPE, is either
continuous or pulse. The calibrator is automatically set to source
mode when OUT is entered. A number and its unit must follow the
command. See Table 7. (Parameter Units) for a list of available units.
For example:

OUT 10 MA sets the calibrator to mA OUT mode and sets the
output to 10mA.

OUT?
Returns the output of the calibrator. Using the above example, OUT?
Would return:

1.000000E-02, A

PRES?
Returns the model and serial number of the attached pressure unit.
Returns NONE if no pressure unit is attached. For example:

PRES? Will return MARTEL,001PNS,3,0

PRES_UNIT?
Returns the pressure units of both the upper and the lower display.
For example if the unit on the upper display is bars, and on the
lower it is psi, the command will return:

BAR, PSI

PULSE_CNT
Sets the number of pulses the calibrator will produce when it is
triggered while in pulse mode. For example;

PULSE_CNT 3000 will set the number of pulses to 3000.