Martel Electronics DMC1410 User Manual

Page 45

Advertising
background image

43

CPRT_COEFA
This command is used for entering a custom RTD into the calibrator.
The numeric value entered after the command will be set as the first
coefficient of the polynomial used by the custom RTD.
For example:

CPRT_COEFA 3.908E-03

enters 3.908e-3 as coefficient A.

CPRT_COEFA?
Returns the number which was entered for the first coefficient of the
polynomial used in the custom RTD. Using the example above
CPRT_COEFA? Would return:

3.908000E-03

CPRT_COEFB
This command is used for entering a custom RTD into the calibrator.
The numeric value entered after the command will be set as the
second coefficient of the polynomial used by the custom RTD.
For example:

CPRT_COEFB -5.8019E-07 enters -5.8019e-7 as coefficient B.

CPRT_COEFB?
Returns the number, which was entered for the first coefficient of the
polynomial used in the custom RTD. Using the example above,
CPRT_COEFB? Would return:

-5.801900E-07

CPRT_COEFC
This command is used for entering a custom RTD into the calibrator.
The numeric value entered after the command will be set as the first
coefficient of the polynomial used by the custom RTD.
For example:

CPRT_COEFC -5.8019E-12 enters -5.8019e-12 as coefficient A.

CPRT_COEFC?
Returns the number which was entered for the first coefficient of the
polynomial used in the custom RTD. Using the example above
CPRT_COEFC? Would return:

-5.801900E-12