Martel Electronics DMC1410 User Manual

Page 55

Advertising
background image

53

ends, the real time clock time stamp is saved, and the technician is
prompted for changes to the tag identification data (any field but tag
number which can not be changed). If test validation was selected,
the technician has the option of displaying the As Found test results.
After either type of As Found test is ended, the technician has the
option of saving the As Found results into the As Left storage and
ending the tag calibration, of entering an adjustment mode to
monitor physical adjustments to the tag prior to starting an As Left
test, or of directly entering the As Left test.
During adjustment mode, the technician can step up and down the
list of test points and the calibrator automatically generates the
corresponding output in the case of a calibrator source type, or
displays the value of a manual output type. In this way any changes
to the original test points resulting from physical adjustments to the
instrument may be evaluated. Alternatively, the technician can enter
output values different from the test points and monitor
corresponding input values. When the adjustment is complete, a
function key is pressed to enter the As Left test.
During the As Left test the calibrator cycles through the test points
one at a time automatically generating the As Found output value in
the case of a calibrator source type, or displaying the As Found
value of a manual output type. Once the input has settled, the
technician presses a function key to prompt for manual input entry if
any, to save both the As Left output and input values, and finally to
step to the next point. Once the last point has been tested, the As
Left test automatically ends and, if test validation was selected, the
technician has the option of displaying the As Left test results.
Finally, the technician has the option of ending the tag calibration, of
entering an adjustment mode to monitor physical adjustments to the
tag prior to starting a further As Left test, or of directly entering a
further As Left test.
Additional As Left tests may be performed after tag calibration has
ended as long as the original tag data has not been cleared from
storage. A menu item is available for selecting any tag and starting
another As Left test
Review functions are available to view test results on the calibrator
display.
Individual or all tag test results may be cleared from storage using
menu functions.

Advertising