Martel Electronics DMC1410 User Manual

Page 34

Advertising
background image

32

2. To enable Remote with Lockout, type in REMOTE and LOCKOUT

in either order.

3. To switch back to local operation enter LOCAL at the terminal.

This command also turns off LOCKOUT if it was on. For more
information on commands refer to the Remote Commands
section.

7.3 Using Commands

7.3-1 Command types
Refer to the Section on Remote Commands for all available
commands.
The calibrator may be controlled using commands and queries. All
commands may be entered using upper or lower case. The
commands are divided into the following categories:
Calibrator Commands
Only the calibrator uses these commands. For example

LOWER_MEAS DCV

tells the calibrator to measure voltage on the lower display.

Common Commands
Standard commands used by most devices. These commands
always begin with an "*". For example

*IDN?

tells the calibrator to return its identification.

Query Commands
Commands that ask for information. They always end with a "?". For
example:

FUNC?

Returns the current modes of the upper and lower displays.

Compound Commands
Commands that contain more than one command on one line. For
example:

LOWER_MEAS RTD; RTD_TYPE CU10

Sets the calibrator to measure RTD in the lower display and sets
RTD type to Cu 10.