100
Table 16: Thermocouple Read and Source (errors in °C)
Range (°C)
Accuracy (°C)
TC Type
Minimum
Maximum
CJC OFF
CJC ON
J
-210.0
-150.0
0.4
0.6
-150.0
1200.0
0.2
0.4
K
-200.0
-100.0
0.5
0.7
-100.0
600.0
0.2
0.4
600.0
1000.0
0.3
0.5
1000.0
1372.0
0.4
0.6
T
-250.0
-200.0
1.5
1.7
-200.0
0.0
0.5
0.7
0.0
400.0
0.2
0.4
E
-250.0
-200.0
1.0
1.2
-200.0
-100.0
0.3
0.5
-100.0
1000.0
0.2
0.4
R
1
0.0
200.0
1.7
1.9
200.0
1767.0
1.0
1.2
S
1
0.0
200.0
1.7
1.9
200.0
1767.0
1.1
1.3
B
1
600.0
800.0
1.5
1.7
800.0
1000.0
1.2
1.4
1000.0
1820.0
1.0
1.2
C
0.0
1000.0
0.5
0.7
1000.0
2316.0
1.5
1.7
XK
-200.0
800.0
0.2
0.4
BP
0.0
800.0
1.9
2.1
800.0
2500.0
0.6
0.8
L
-200.0
900.0
0.2
0.4
U
-200.0
0.0
0.4
0.6
0.0
600.0
0.2
0.4
N
-200.0
-100.0
0.8
1.0
-100.0
1300.0
0.3
0.5
CJC error outside of 23 ± 5°C is 0.05°C/°C
1. For thermocouple measure mode on B, R and S; round the specification up or
down accordingly as there is no resolution past the decimal point.