Table 16: Thermocouple Read and Source (errors in °C)

Range (°C)

Accuracy (°C)

TC Type

Minimum

Maximum

CJC OFF

CJC ON

J

-210.0

-150.0

0.4

0.6

-150.0

1200.0

0.2

0.4

K

-200.0

-100.0

0.5

0.7

-100.0

600.0

0.2

0.4

600.0

1000.0

0.3

0.5

1000.0

1372.0

0.4

0.6

T

-250.0

-200.0

1.5

1.7

-200.0

0.0

0.5

0.7

0.0

400.0

0.2

0.4

E

-250.0

-200.0

1.0

1.2

-200.0

-100.0

0.3

0.5

-100.0

1000.0

0.2

0.4

R

1

0.0

200.0

1.7

1.9

200.0

1767.0

1.0

1.2

S

1

0.0

200.0

1.7

1.9

200.0

1767.0

1.1

1.3

B

1

600.0

800.0

1.5

1.7

800.0

1000.0

1.2

1.4

1000.0

1820.0

1.0

1.2

C

0.0

1000.0

0.5

0.7

1000.0

2316.0

1.5

1.7

XK

-200.0

800.0

0.2

0.4

BP

0.0

800.0

1.9

2.1

800.0

2500.0

0.6

0.8

L

-200.0

900.0

0.2

0.4

U

-200.0

0.0

0.4

0.6

0.0

600.0

0.2

0.4

N

-200.0

-100.0

0.8

1.0

-100.0

1300.0

0.3

0.5


CJC error outside of 23 ± 5°C is 0.05°C/°C