99 table 12: frequency measurement/source, Table 13: resistance measurement, Table 14: resistance source – Martel Electronics DMC1410 User Manual

Page 101: Table 15: thermocouple measurement/source

Advertising
background image

99

Table 12: Frequency Measurement/Source

Range

Accuracy

(% of reading ± floor)

Read

2.0CPM - 600.0CPM

0.05% ± 0.1CPM

1.0Hz - 1000.0Hz

0.05% ± 0.1Hz

1.00KHz - 10.00KHz

0.05% ± 0.01KHz

Source

2.0CPM - 600.0CPM

0.05%

1.0Hz - 1000.0Hz

0.05%

1.00KHz - 10.00KHz

0.250%

Input voltage amplitude range on frequency is 1V to 20V zero based square wave only.Output amplitude

is adjustable from 1V to 20V, and is a square wave with 50% duty cycle.For output frequency, a slight
negative offset of approximately -0.1V is present to assure zero crossing.

Table 13: Resistance Measurement

Range

Accuracy

(% of reading ± floor)

Ohms low

0.00Ω - 400.0Ω

0.015% ± 0.03Ω

Ohms high

401.0Ω - 4000.0Ω

0.015% ± 0.3Ω

Table 14: Resistance Source

Range

Excitation Current

Accuracy

(% of reading ± floor)

Ohms low 5.0Ω - 400.0Ω

0.1mA - 0.5mA

0.015% ± 0.1Ω

5.0Ω - 400.0Ω

0.5mA - 3mA

0.015% ± 0.03Ω

Ohms high 400Ω - 1500Ω

0.05mA - 0.8mA

0.015% ± 0.3Ω

1500Ω - 4000Ω

0.05mA - 0.4mA

0.015% ± 0.3Ω

Note: Unit is compatible with smart transmitters and PLCs. Frequency response is <= 5ms.

Table 15: Thermocouple Measurement/Source

Range

Accuracy

(% of reading ± floor)

Read (mV)

-10.000mV - 75.000mV

0.015% ± 10µV

Source (mV)

-10.000mV - 75.000mV

0.015% ± 10µV

Maximum current output in voltage ranges is 1mA with an output impedance of <= 1Ω