110

An unknown command was received

111

An invalid RTD or TC parameter value was received

112

The serial input buffer overflowed

113

Too many entries in the command line

114

The serial output buffer overflowed

115

Output is overloaded

116

Calibrator not in pulse train mode when TRIG was received

117

An invalid FREQ_TYPE was received

7.5 Entering Commands

Commands for the calibrator may be entered in upper or lower case.
There is at least one space required between the command and
parameter, all other spaces are optional. Almost all commands for
the calibrator are sequential, any overlapped commands will be
indicated as such. This section will briefly explain each of the
commands and describe their general use, which will include any
parameters that may be entered with the command as well as what
the output of the command is.

7.5-1 Common Commands
*CLS
Clears the ESR, the error queue and the RQS bit. Also terminates all
pending operations. When writing programs, use before each
procedure to avoid buffer overflow.

*ESE
Loads a byte into the Event Status Enable register. The command is
entered with a decimal number that, when converted to binary,
enables the right bits in the Event Status Register. For example:

*ESE 133

When 133 is converted to binary it is 10000101. Bits 7, 2, and 0 will
be enabled.

*ESE?
Returns the contents of the Event Status Enable register. The value
returned is a decimal. For example, if the register has the following
settings:

10000101

than the value returned will be 133.

40