Martel Electronics DMC1410 User Manual

Page 41

Advertising
background image

Table 7: Parameter units

Units

Meaning

MA

milliamps of current

MV

Voltage in millivolts

V

Voltage in volts

CPM

Frequency in cycles per minute

Hz

Frequency in Hertz

KHz

Frequency in kiloHertz

Ohms

Resistance in Ohms

Cel

Temperature in Celsius

Far

Temperature in Fahrenheit

Psi

Pressure in pounds per square-inch

InH2O4C

Pressure in inches of water at 4°C

InH2O20C

Pressure in inches of water at 20°C

CmH2O4C

Pressure in centimeters of water at 4°C

CmH2O20C

Pressure in centimeters of water at 20°C

Bar

Pressure in bars

Mbar

Pressure in millibars

KPal

Pressure in kiloPascals

InHg

Pressure in inches of mercury at 0°C

MmHg

Pressure in millimeters of mercury at 0°C

Kg/cm2

Pressure in kilograms per square-centimeter

Table 8: Error codes

Error Number

Error Description

100

A non-numeric entry was received where it should be a
numeric entry

101

Too many significant digits entered

102

Invalid units or parameter value received

103

Entry is above the upper limit of the allowable range

104

Entry is below the lower limit of the allowable range

105

A required command parameter was missing

106

An invalid pressure unit was received

107

An invalid CJC_STATE was received

108

An invalid TSENS_TYPE was received

109

Pressure module not connected

39