13 analysis method: baseline correction, 1 assay requirements, 2 setup – Techne PrimeQ User Manual

Page 114

Advertising
background image

114

3.13 Analysis method: Baseline correction

This method allows the user to adjust the data for any background fluorescence. This approach
looks at the signal in the early cycles of the PCR and then averages out the early noise and
subtracts it from subsequent readings. As with passive reference dye (PRD) correction detailed
above, baseline correction can be useful for correcting the data prior to performing an analysis and
so can help to increase the accuracy of the assay.

3.13.1

Assay requirements

• Need at least one reporter.

• Need at least two readings.

• Need a dye defined as a PRD for PRD correction options to be available.

3.13.2

Setup

The diagram below gives an overview of the different baseline correction methods.

• Select Baseline from the drop-down menu in the Analysis Wizard Selection box. Assign the

dye usage and click Next. The Baseline Correction Wizard will launch.

3.13.2.1 PRD correction

If a PRD was assigned in the dye usage screen, the first screen of the wizard will ask if the data is
to be PRD corrected.

Raw data for Reporter with

or without PRD correction

None:
No baseline
correction

Arithmetic 1:
1.

For each well, calculate
average of readings in
range x to y.

2.

Subtract average from
all readings

Arithmetic 2:
1.

For each well, calculate
average of n lowest
readings.

2.

Subtract average from all
readings.

Proportional:
1.

For each well, calculate
average of n1 lowest
readings.

2.

Calculate average of n2
highest readings
(maximum).

3.

Calculate each reading as
% of the maximum.

SELECT BASELINE CORRECTION METHOD

Software plots fluorescence vs.

cycle no. using corrected data

Raw data for Reporter with

or without PRD correction

None:
No baseline
correction

Arithmetic 1:
1.

For each well, calculate
average of readings in
range x to y.

2.

Subtract average from
all readings

Arithmetic 2:
1.

For each well, calculate
average of n lowest
readings.

2.

Subtract average from all
readings.

Proportional:
1.

For each well, calculate
average of n1 lowest
readings.

2.

Calculate average of n2
highest readings
(maximum).

3.

Calculate each reading as
% of the maximum.

SELECT BASELINE CORRECTION METHOD

Software plots fluorescence vs.

cycle no. using corrected data

Advertising