1 assay requirements, 2 setup – Techne PrimeQ User Manual

Page 135

Advertising
background image

135

3.15.1

Assay requirements

• Need at least one reporter dye.

• Need a stage in the program with a ramp read (with at least two readings).

3.15.2

Setup

• Program a ramp read into the PCR program as described in section 3.4.3.6.

• In the Analysis Selection box in the Experiment Editor, click on the name of the ramp stage

and then press

Edit.

• This will lead through to the Analysis Wizard Selection box. Select Dissociation Curve in

the drop-down menu and select dye usage.

• Click Next and the Dissociation Curve Wizard will launch.

The wizard will lead the user through the setup:

Background correction: Corrects for effects of temperature on fluorescence and the

background before and after all the DNA has melted.

Digital filter: Smoothes the dissociation curve raw data (but does not affect the calculated

results).

Peak detector: Finds specified number of peaks and records the dissociation temperature

(Tm).

Peak area: Calculates the area under the peaks detected.

3.15.2.1 Background correction

The first window of the wizard allows setting of background correction parameters.

Raw data

for reporter

Background

correction and

optional curve

smoothing filter

Dissociation

curve

Dissociation

peaks

Report

Tm

Peak areas

Raw data

for reporter

Background

correction and

optional curve

smoothing filter

Dissociation

curve

Dissociation

peaks

Report

Tm

Peak areas