A message will appear informing the user to

Please wait – checking cartridge status.

If the removal was performed correctly, the Cartridge Access and Editing table will be updated to
display “Not Fitted” next to the slot from which the cartridge was removed.

The table below gives details of the PrimeQ filter cartridges.

Filter

cartridge

Excitation

wavelength (nm)

Emission

wavelength (nm)

Suitable dyes

*FC01

460

500

FAM™ multiplex, Green
intercalating dyes, Fluorescein,
EvaGreen

®

FC02

485

520

FAM™, Green intercalating,
Fluorescein, EvaGreen

®

FC03

530

560

HEX™, TET™, JOE™, VIC

®

,

Yakima Yellow

®

FC04

580

615

ROX™, TEXAS RED

®

, Cy

®

3.5

FC05

640

685

Cy

®

5, Quasar

®

670

*FC01 may be ordered separately.

1.16.8

Points to remember

• Filter descriptions are stored in the instrument memory and do not need to be re-entered if

the connected instrument is changed to a different PC. Filter descriptions are uploaded from
the instrument when Quansoft is opened.

• Changing the filter cartridge between experiments can only be carried out via the software –

it cannot be performed manually.

• It is critically important that the correct excitation and emission filters are in the assigned

position of the carousel.

• During multiplexing you must choose two fluorophores with different excitation and emission

wavelengths and measure the fluorescence with different filter cartridges.

• Avoid combinations of fluorophores with overlapping spectra.

• It is possible for some fluorophores to appear more than once on the dye list because more

than one filter pair could be used for its measurement. In this case, select the correct filter
by its wavelength properties.

1.17 Changing the instrument name

All PrimeQ units are shipped with the name “PrimeQ” stored in the instrument’s software. This can
be changed to a more appropriate name as required.

• To change the instrument name, click on the Maintenance icon on the navigation bar. Then

click on the

Instruments icon. The default supervisor password is techne – we suggest you

change this as soon as possible by clicking on the

Security icon. Please ensure that you

keep a record of the password, as without it you will not be able to access these functions.