Undo: Cancels the previous operation.

Rotate plate 180°: Useful if the plate was loaded into the instrument the opposite
way round to the designated layout.

These functions can also be accessed from the Tools option in the menu bar.

3.4.4.7

Well information table

As a sample is allocated to each well,
information about the sample will appear in the
adjacent table. A default name and number is
assigned e.g. Standard 1, but this can easily be
changed by highlighting and typing over the
text.

Use the Reset Names function to return any
user-defined

changes

to

the

default.

Concentrations can be added for standards
(units defined by the drop-down menu) and any
comments added.

Choose

to

add

user-defined

names,

concentrations and units of standards or any
comments specific to a sample.

The Well Information table will contain a complete list of all the sample types in the plate, but to list
the details of just a selected sample type, click the Selected Sample Type button and then click
on a sample type icon to choose which sample information to display. For instance, to display only
the well information for the standards, click on the Selected Sample Type button and then click on
the red STD sample icon above the plate layout. When only the selected well information is being
shown, the function button will now display Select All Wells. Click on this button to return to
displaying the complete well information.

• Press Done when complete. Quansoft will return to the Experiment Editor and the new plate

layout will appear in the plate layout pane. Press Cancel at any time to abort the procedure
and return to the Experiment Editor.