Troubleshooting – BrandTech Scientific HandyStep electronic User Manual

Page 36

Advertising
background image

- 36 -

Problem

Error message displayed

Possible cause

Corrective action

Tip dripping

No message

Leaking tip

Replace tip

Faint display;

Battery symbol flashing

Battery low

Recharge battery pack

reduced speed

or replace with charged battery pack.

Instrument not responding

ERR 001

- Locking lever not engaged

1. Press ENTER to confirm error.

during aspirating or dispensing

Follow MENU-instructions to avoid instrument damage!

- Locking lever open at lower

Display: CLO (Close)

third of the tip

2. Engage the locking lever

Display: rEF (Reference run)

3. Press the STEP key

Attention: Liquid remaining in tip will be discharged

during the reference run (rEF). Discharge sample into

a suitable vessel (see pages 5, point 10).

Instrument not responding

ERR 002

Motor blocked

1. Press ENTER to confirm error.

Follow MENU-instructions to avoid instrument damage!

- Battery low

- Viscosity is too high (see page 7) - Recharge or replace battery pack

- Tip orifice obstructed

- Reduce aspirating/dispensing speed

Display: rEF (Reference run)

2. Press the STEP key

Attention: Liquid remaining in tip will be discharged

during the reference run (rEF). Discharge sample into

a suitable vessel (see pages 5, point 10).

Instrument not responding

ERR TIP

Locking lever opened too early

1. Press ENTER to confirm error.

within the lower third of the tip

Follow MENU-instructions to avoid instrument damage!

2. Display: OPE (Open)

Push the locking lever to its upper stop.

Troubleshooting