Troubleshooting – BrandTech Scientific HandyStep electronic User Manual

Page 37

Advertising
background image

- 37 -

Problem

Error message displayed

Possible cause

Corrective action

Instrument not responding

3. Display alternating between out TIP and CLO.

- continued -

Remove tip and close locking lever (without tip).

4. HandyStep

®

electronic automatically carries out a

reference run.

5. Display: no TIP

Insert new tip (see page 16).

Instrument not responding

ERR 003

Electronic fault or tip connector

1. Press ENTER to confirm error.

not at lowest position when

2. Follow MENU-instructions:

inserting new tip

- Display rEF (reference run)

- Press STEP-Key

Attention: Liquid remaining in tip will be discharged

during reference run (rEF). Discharge sample into

suitable vessel (see page 5, point 10).

If error message is still shown on the display remove

battery pack and insert it again (reset).

If this does not solve the problem send instrument back for

repair.

No tone, no light when

ACC

- Instrument contains no battery

- Insert battery pack.

placing the HandyStep

®

- Battery pack inserted improperly - Make sure battery contacts are aligned with contacts

electronic into the

in the instrument.

charging dock

- Battery pack faulty

- Replace battery pack.

Charge-indicator

No message

Removal or insertion of

- Charge battery pack for at least 3.5 hours

is flashing

a new battery pack

inside the HandyStep

®

electronic.

Dispensing process

ERR VOL

Dispensing process interrupted by

interrupted.

Attention!

- Pressing the

∆ key during

1. Press ENTER to confirm error.

Volume not completely

dispensing

discharged

- electronic faults

2. Restart dispensing by pressing the STEP key.

Troubleshooting

- continued -

Note:
If other error messages are displayed, remove battery pack and insert it again.
If this does not solve the problem, send instrument in for repair.