Autozero function, Utozero function – RADWAG TMX 6/15/H2 Multifunctional Scales User Manual

Page 105

Advertising
background image


After selecting the required value save changes according to 29.3.2
of the manual.



29.6. Autozero function

The autozero function has been introduced in order to assure precise
scale’s indications. This function controls and corrects „0” indication.
While the function is active it compares the results successively with
constant frequency. If results differ less than the declared value of autozero
range, for example 1unit, the scale will be automatically zeroed and the
stability result markers

and zero indicator

will be displayed.

When AUTOZERO function is active, each measurement starts with
precise zero. In particular cases the function can disturb measuring
process. For example when product is loaded very slowly on scale
pan (pouring), Autozero system can influence real weighing results.


Procedure:

• Readout parameters value according to 29.3.1 of the manual,
• Press, then rewind Autozero parameter window.

- 105 -