Closing weighingserver program, Error messages, Label designing – RADWAG TMX 6/15/H2 Multifunctional Scales User Manual
Page 107
Notice:
Additional weighing platform module MW02
is working on port COM2.
Languages available:
• Auto
-
Setting language version compatible with the
version of the operating system.
• Polish
• English
• German
Set the address track to the software keyboard on:
C:\Program Files\screen keyboard\KEYBOARD_mini.exe
29.8. Closing WeighingServer program
In order to exit the program press the button:
on the main
window.
30. ERROR MESSAGES
Err2
- Value beyond zero range
Err3
- Value beyond tare range
Err8
- Exceeded time of tarring / zeroing operation
null
- Converter zero value
FULL
- Measurement range exceeding
LH
- Initial mass error, indication beyond
(from -5% to +15% range of initial mass)
31. LABEL DESIGNING
Computer program EDYTOR ETYKIET R01 is used for label patterns
making. First a label pattern is saved as a file with “lb” extension, then
it is set in Terminal E2R Ewidencja program.
- 107 -