34.2.1. Technical specification

Parameters of outputs

Quantity of outputs

8

Type of outputs

Reed operation contacts

Wire diameter

0,14 - 0,5mm

2

Maximal load-current contact capacity

0,2A DC

Maximal forward voltage

50V DC

Parameters of inputs

Quantity of inputs

8

Input type

Optoinsulated

Wire diameter

0,14 – 0,5mm

2

Control voltage range

5 -24V DC



34.2.2. Installing method in PUE 5 terminal

Module is designated for assembly inside terminal PUE5. Module
is assembled to main board of terminal to 10-pin J22 connector. For
module WE8 gland is installed on casing lid, 3m cable ended with
insulated conductors is led through gland.


Installing procedure:

1. Unplug the terminal (remove the plug from socket 230V);
2. Unscrew the casing lid (back part of terminal casing);
3. Install the module in connector J22 on main board;
4. While installing module pay attention to plastic posts fastening

to main board. They should be placed in assembly holes in main
board and in assembly holes in module WE 8,

5. In one of free glands remove the stopper and led through it cable

IN/OUT (if necessary use bigger gland depending on cable diameter);

6. Connect the cable IN/OUT to joint J1, J2 for outputs and J3, J4 for

inputs on module 8IN/8OUT according to description given in table;

7. Connect cable IN/OUT to group of conductors (undo band clips

fastening group of conductors, lay the cable and clamp band clips –
multiple use band clips);

8. Assembly cover of indicator casing.