cAlibrAtion

gT3000 wITH TOxIC gAS SENSOR

From gT3000:
1. Using the magnet, activate the magnetic calibration

switch on the GT3000. The green LED turns to

yellow.

MAGNETIC

SWITCH

LED

A2443

location of Magnetic Switch on GT3000

2. The UD10 will display “Waiting for Zero” on the main

display screen, with the yellow LED on the detector

housing illuminated continuously.

3. The UD10 will then display “Waiting for Signal” on

the screen, while the yellow LED on the detector

is flashing. The device automatically performs the

zero calibration.

4. The UD10 will display “Waiting for Gas” on the

screen, while the yellow LED on the detector is

flashing.

5. Apply calibration gas to the sensor.

6. The UD10 will display “Waiting for Span” on the

screen, while the yellow LED on the detector is

flashing.

7. When the UD10 displays “Remove Cal Gas” on the

screen and all LEDs on the detector housing are off,

remove the calibration gas.

8. After successful calibration, the UD10 automatically

returns to the normal display with the green LED

illuminated on the detector.

From UD10
1. Using the magnet to activate the switches on the

UD10 display, navigate to the Calibration menu.

Main Menu
Process Vars
Display Status
Device Status
Display Setup
Device Setup
Device Cal
Display Test
Device Test

Device Cal

Cal Gas Conc
Calibration
Change Snsr Type

Calibration

Execute
Abort

2. Activate “Execute” (Enter/Select) to start calibration.

3. The UD10 will display “Waiting for Zero” on the

main display screen, with the yellow LED on the

detector housing illuminated continuously. The

device automatically performs the zero calibration.

4. The UD10 will then display “Waiting for Gas” on

the screen, while the yellow LED on the detector is

flashing.

5. Apply calibration gas to the sensor.

6. The UD10 will display “Waiting for Span” on the

screen, while the yellow LED on the detector is

flashing.

7. When the UD10 displays “Remove Cal Gas” on the

screen and the LEDs on the detector housing are

off, remove the calibration gas.

8. After completion of a successful calibration, the

UD10 automatically returns to the normal mode with

the green LED illuminated on the detector.

UD10 mA Output During Calibration

(UD10 with gT3000)

UD10 Display Reading

Standard

mode

Replicate

mode

Waiting for Zero

3.8

3.8

Waiting for Gas

3.8

3.8

Waiting for Span

3.8

3.8

Remove Cal Gas

3.8

3.8

back to normal

4.0