Calibration – Det-Tronics UD10 FlexVu Explosion-Proof Universal Display Unit User Manual

Page 45

Advertising
background image

i-3

95-8661

4.2

cAlibrAtion

To initiate calibration of the PIRECL from the UD10

Display:

1. Using the magnet to activate the switches on the

UD10 display, navigate to the “Calibration” menu.

Main Menu
Process Vars
Display Status
Device Status
Display Setup
Device Setup
Device Cal
Display Test
Device Test

Device Cal

Cal Gas Con
Cal Gas Type
Calibration
Date

Calibration

Execute
Abort

2. Activate “Execute” (Enter/Select) to start calibration.

3. The UD10 will display “Waiting for Zero” on the main

display screen, with a solid red LED on the PIRECL

housing illuminated.

4. The UD10 will then display “Waiting for Gas” on the

screen, while the LED on the PIRECL is flashing red.

5. Apply calibration gas to the PIRECL

6. The UD10 will display “Waiting for Span” on the

screen, while a red flashing LED on the PIRECL

housing is illuminated.

7. When the UD10 displays “Remove Cal Gas” on the

screen and the LED on the PIRECL housing is off,

remove the calibration gas.

8. After successful calibration, the UD10 automatically

returns to the normal display with the green LED

illuminated on the PIRECL housing.

To initiate calibration from the PIRECL while monitoring

calibration using the UD10 display:

1. Using the magnet, activate the magnetic calibration

switch on the PIRECL detector. See Figure below.

The LED turns from green to red.

2. The UD10 will display “Waiting for Zero” on the main

display screen, with a solid red LED on the PIRECL

housing illuminated.

3. The UD10 will then display “Waiting for Gas” on the

screen, while the LED on the PIRECL is flashing red.

4. Apply calibration gas to the PIRECL detector.

5. The UD10 will display “Waiting for Span” on the

screen, while a red flashing LED on the PIRECL

housing is illuminated.

6. When the UD10 displays “Remove Cal Gas” on the

screen and the LED on the PIRECL housing is off,

remove the calibration gas.

7. After successful calibration, the UD10 automatically

returns to the normal mode with the green LED

illuminated on the PIRECL.

CALIBRATION MAGNET

CALIBRATION NOZZLE

MULTICOLOR LED

HART COMMUNICATION PORT

(COVER INSTALLED)

PLACE CALIBRATION MAGNET
HERE TO ACTIVATE INTERNAL
REED SWITCH

B2435

Model PiRECl Gas Detector

UD10 mA Output During Calibration

(UD10 with pIRECl)

UD10 Display Reading

Standard

mode

Replicate

mode

Waiting for Zero

3.8

2.2

Waiting for Gas

3.8

2.0

Waiting for Span

3.8

2.0

Remove Cal Gas

3.8

1.8

back to normal

4.0

4.0